The HI-8733 portable conductivity meter is designed to measure conductivity in water and solutions for water quality control in water treatment processes, environmental studies, swimming pools, etc.
- With manual calibration and automatic temperature compensation.
- Supplied with a four-electrode conductivity sensor that allows operation in four measuring ranges.
- Can be calibrated at one point using a standard conductivity solution. Using the calibration wheel on the top of the instrument, the instrument can be easily calibrated to a calibration standard.
| Name | Meaning |
| Conductivity range | from 0.0 to 199.9 µS/cm; from 0 to 1999 µS/cm; from 0.00 to 19.99 mS/cm; 0.0 to 199.9 mS/cm |
| Conductivity resolution | 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm |
| Conductivity Accuracy | ±1% of full scale (excluding sensor error) |
| Conductivity calibration | manual, one point at a time using the calibration wheel |
| temperature compensation | automatic, with β adjustable from 0.0 to 2.4% / ° C temperature range 0.0 to 50.0 °C / 32.0 to 122.0 °F |
| Electrode probe | HI76302W conductivity probe with built-in temperature probe, DIN connector and 1 m (3.3') cable (included) |
| Battery Type/Durability | 9 V / approx. 100 hours of continuous operation |
| Environment | 0 to 50°C (32 to 122 F); relative humidity max 95% |
| Dimensions | 145x80x36 mm (5.7x3.1x1.4”) |
| The weight | 230 g |
The HI8733 is supplied with the HI76302W Conductivity Probe, sachets of HI 70030 12880 µS/cm Calibration Solution (5), battery, instruction manual, and rugged carrying case.